FDA Device recall Z-1958-2012
Central Control Module for System 1: System 1 Base, 220/240V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours…
- FDA Device
- Class II
- terminated
- Published 2012-07-18
On 2012-07-18 the FDA classified a Class II recall of Central Control Module for System 1: System 1 Base, 220/240V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours… by Terumo Cardiovascular Systems, citing terumo CVS has received reports of the Central Control Monitor (CCM) for Terumo Advanced Perfusion System 1 losing partial or full functionality. Symptoms that have been associate….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1958-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-06-22 |
| Published | 2012-07-18 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | US |
Product
Central Control Module for System 1: System 1 Base, 220/240V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Reason
Terumo CVS has received reports of the Central Control Monitor (CCM) for Terumo Advanced Perfusion System 1 losing partial or full functionality. Symptoms that have been associated with a malfunctioning CCM include: full or partial loss of display, loss of control functions; inability to distinguish various status, alert, alarm conditions, the inability to operate the touch screen, or the CCM
Identifiers
| code_info | Catalog number: 801764 and serial numbers: 0006-0066, 0100-0311, 0313- 0322,1001-1027, 1100-1434, 1436, 1440, and 1441. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1958-2012%22