FDA Device recall Z-1957-2025

KWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived from ATCC¿ 17960

On 2025-06-18 the FDA classified a Class II recall of KWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived from ATCC¿ 17960 by Microbiologics, citing affected products were packaged with the incorrect KWIK STIK assembly resulting in the product containing the wrong microorganism. 0621K was packaged 01065K swabs; 01065K was pack….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1957-2025
ClassificationClass II
Statusongoing
Initiated2025-05-27
Published2025-06-18
Last updated2026-09-13 11:57 UTC
CompanyMicrobiologics
DistributionUS

Product

KWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived from ATCC¿ 17960; 0621K, Staphylococcus aureus derived from NCTC 12493

Reason

Affected products were packaged with the incorrect KWIK STIK assembly resulting in the product containing the wrong microorganism. 0621K was packaged 01065K swabs; 01065K was packaged with 0621K swabs. This may lead to delayed diagnosis.

Identifiers

code_info01065K: UDI-DI 30845357032868, lot 1065-25-51; 0621K: UDI-DI 30845357014666, lot 621-73-21