FDA Device recall Z-1957-2021

A-Series BiPAP Hybrid A30 (not marketed in US), A-Series BiPAP V30 Auto Continuous Ventilator, Minimum Ventilatory Support, Facility Use

On 2021-07-21 the FDA classified a Class I recall of A-Series BiPAP Hybrid A30 (not marketed in US), A-Series BiPAP V30 Auto Continuous Ventilator, Minimum Ventilatory Support, Facility Use by Philips Respironics, citing the polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may o….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1957-2021
ClassificationClass I
Statusongoing
Initiated2021-06-14
Published2021-07-21
Last updated2026-09-13 11:56 UTC
CompanyPhilips Respironics
DistributionUS

Product

A-Series BiPAP Hybrid A30 (not marketed in US), A-Series BiPAP V30 Auto Continuous Ventilator, Minimum Ventilatory Support, Facility Use

Reason

The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.

Identifiers

code_infoAll Devices manufactured before 26 April 2021, All device serial numbers. Material Numbers 1111181 1111182 1111178 1140184 1135427 1143671

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1957-2021%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.