FDA Device recall Z-1956-2026
Philips Ingenia 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781315. 2. Model Number (REF): 781341. 3. Model Number (REF): 78139…
- FDA Device
- Class II
- ongoing
- Published 2026-05-06
On 2026-05-06 the FDA classified a Class II recall of Philips Ingenia 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781315. 2. Model Number (REF): 781341. 3. Model Number (REF): 78139… by Philips North America, citing the potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1956-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-04-14 |
| Published | 2026-05-06 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Philips North America |
| Distribution | US |
Product
Philips Ingenia 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781315. 2. Model Number (REF): 781341. 3. Model Number (REF): 781396. 4. Model Number (REF): 782115. 510(k) Numbers: K163116, K173079, K183063, K193215, K213583, K223458, K251397, K251808
Reason
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Identifiers
| code_info | 1. Model Number (REF): 781315. UDI: (01)00884838099715(21)61067, (01)00884838099715(21)61058, (01)00884838099715(21)610…1. Model Number (REF): 781315. UDI: (01)00884838099715(21)61067, (01)00884838099715(21)61058, (01)00884838099715(21)61056. Serial Numbers: 61067 61058, 61056. 2. Model Number (REF): 781341. UDI: (01)00884838055322(21)70229, (01)00884838055322(21)84667, (01)00884838055322(21)70311, (01)00884838055322(21)70668, (01)00884838055322(21)70845, (01)00884838055322(21)70851, (01)00884838055322(21)87151, (01)00884838055322(21)87112, (01)00884838055322(21)84322, (01)00884838055322(21)84323, (01)00884838055322(21)84228, (01)00884838055322(21)84789, (01)00884838055322(21)84261, (01)00884838055322(21)84500, (01)00884838055322(21)84556, (01)00884838055322(21)87187, (01)00884838055322(21)84840, (01)00884838055322(21)70685, (01)00884838055322(21)84661, (01)00884838055322(21)70787, (01)00884838055322(21)70063, (01)00884838055322(21)70621, (01)00884838055322(21)70743, (01)00884838055322(21)70911, (01)00884838055322(21)84049, (01)00884838055322(21)84088, (01)00884838055322(21)84147, (01)00884838055322(21 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1956-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.