RecallRadar

FDA Device recall Z-1952-2013

Synthes Flexible Medullary Reamer

On 2013-08-21 the FDA classified a Class II recall of Synthes Flexible Medullary Reamer by Synthes Usa Hq, citing due to the coiled design for this product, the product is difficult to clean and the potential for corrosion on the device exists. The potential for harm exists if there is an int….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1952-2013
ClassificationClass II
Statusterminated
Initiated2013-06-11
Published2013-08-21
Last updated2026-09-13 11:53 UTC
CompanySynthes Usa Hq
DistributionUS

Product

Synthes Flexible Medullary Reamer. Intended to be used to facilitate the preparation of the intramedullary cavity.

Reason

Due to the coiled design for this product, the product is difficult to clean and the potential for corrosion on the device exists. The potential for harm exists if there is an introduction of foreign material at the operative site originating from the presence of corrosion and/or the retained undefined material that could not be adequately cleaned.

Identifiers

code_infoAll lots with part number 359.106, 359.107, 359.108, 359.109, 359.110, 359.111, 359.112, 359.113, 359.114, 359.115

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1952-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.