On 2013-08-21 the FDA classified a Class II recall of Synthes Flexible Medullary Reamer by Synthes Usa Hq, citing due to the coiled design for this product, the product is difficult to clean and the potential for corrosion on the device exists. The potential for harm exists if there is an int….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1952-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-06-11 |
| Published | 2013-08-21 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Synthes Usa Hq |
| Distribution | US |
Product
Synthes Flexible Medullary Reamer. Intended to be used to facilitate the preparation of the intramedullary cavity.
Reason
Due to the coiled design for this product, the product is difficult to clean and the potential for corrosion on the device exists. The potential for harm exists if there is an introduction of foreign material at the operative site originating from the presence of corrosion and/or the retained undefined material that could not be adequately cleaned.
Identifiers
| code_info | All lots with part number 359.106, 359.107, 359.108, 359.109, 359.110, 359.111, 359.112, 359.113, 359.114, 359.115 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Synthes Usa Hq
- Synthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicated for screw insertion, pin and w…
FDA Device · Class II · 2014-01-08 · Synthes Usa Hq - Synthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFN Nail is intended to treat stable…
FDA Device · Class II · 2013-12-18 · Synthes Usa Hq - Synthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedge osteotomies of the medial proximal ti…
FDA Device · Class II · 2013-12-18 · Synthes Usa Hq - Synthes Soft Tissue Retractor Small Extendible
FDA Device · Class II · 2013-12-11 · Synthes Usa Hq - Titanium Trochanteric Fixation Nail (TFN)System Intended to treat stable and unstable fractures of the proximal femur
FDA Device · Class II · 2013-12-11 · Synthes Usa Hq
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1952-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.