FDA Device recall Z-1952-2013
Synthes Flexible Medullary Reamer
- FDA Device
- Class II
- terminated
- Published 2013-08-21
On 2013-08-21 the FDA classified a Class II recall of Synthes Flexible Medullary Reamer by Synthes Usa Hq, citing due to the coiled design for this product, the product is difficult to clean and the potential for corrosion on the device exists. The potential for harm exists if there is an int….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1952-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-06-11 |
| Published | 2013-08-21 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Synthes Usa Hq |
| Distribution | US |
Product
Synthes Flexible Medullary Reamer. Intended to be used to facilitate the preparation of the intramedullary cavity.
Reason
Due to the coiled design for this product, the product is difficult to clean and the potential for corrosion on the device exists. The potential for harm exists if there is an introduction of foreign material at the operative site originating from the presence of corrosion and/or the retained undefined material that could not be adequately cleaned.
Identifiers
| code_info | All lots with part number 359.106, 359.107, 359.108, 359.109, 359.110, 359.111, 359.112, 359.113, 359.114, 359.115 |
|---|