FDA Device recall Z-1948-2014

Manual Resuscitator Bag Series: AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM1000, CPRM2000, CPRM300…

On 2014-07-09 the FDA classified a Class I recall of Manual Resuscitator Bag Series: AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM1000, CPRM2000, CPRM300… by Ventlab, citing the pop-off valve in the defective devices remain open and a squeeze of the bag may not generate enough force to force the duck bill valve open and therefore no air will be delive….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1948-2014
ClassificationClass I
Statusterminated
Initiated2014-05-16
Published2014-07-09
Last updated2026-09-13 11:53 UTC
CompanyVentlab
DistributionUS

Product

Manual Resuscitator Bag Series: AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM1000, CPRM2000, CPRM3000, PRO-1900, PRO-1000, PRO-2000, SC7000, SC8000, SC8020, SC8120, SC9000, SS3200, VN2000, VN3000, VN4000, VN5000, VT1000 , VN2102, VN2002. Emergency manual resuscitator for the Pediatric Population.

Reason

The pop-off valve in the defective devices remain open and a squeeze of the bag may not generate enough force to force the duck bill valve open and therefore no air will be delivered to the victim.

Identifiers

code_info
Manual Resuscitator Bag Series: AF1000, AF2000, AF5000, BT4000, VN2000, VN5000, VT1000; Lot #s105147 -107609. AF1000…Manual Resuscitator Bag Series: AF1000, AF2000, AF5000, BT4000, VN2000, VN5000, VT1000; Lot #s105147 -107609. AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, PRO-1900, SC7000, SC8120, SS3200, VN2000, VN3000, VN4000, VN5000; Lot #s 106245 -107291. VN2102; Lot #200349. AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM2000, CPRM3000, PRO-1000, PRO-2000, SC7000, SC8020, VN2000, VN3000, VN4000; Lot #s 107029 -107634. VN2002; Lot # 200492. SC8000, SC9000; Lot #s 101441 -107461. CPRM1000; Lot #s 99523 -107315.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1948-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.