FDA Device recall Z-1946-2024
Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone
- FDA Device
- Class II
- ongoing
- Published 2024-06-12
On 2024-06-12 the FDA classified a Class II recall of Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone by Newport, citing according to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1946-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-03-27 |
| Published | 2024-06-12 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Newport |
| Distribution | US |
Product
Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone.
Reason
According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.
Identifiers
| code_info | P/N N-LHP-928 Set S/N (laser head + power supply): 6905AE, 6907AE, 9023AE, 3148AF, 6899AE 6909AE, 9025AE, 5695AF, 5696A…P/N N-LHP-928 Set S/N (laser head + power supply): 6905AE, 6907AE, 9023AE, 3148AF, 6899AE 6909AE, 9025AE, 5695AF, 5696AF, 5691AF, 5692AF, 5693AF, 5694AF (9 of them in Irvine warehouse) P/N: N6-LPL-944-080 Power supply S/N: 23060081, 23060079 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1946-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.