RecallRadar

FDA Device recall Z-1944-2020

iChem¿VELOCITY Urine Chemistry System-

On 2020-05-20 the FDA classified a Class II recall of iChem¿VELOCITY Urine Chemistry System- by Beckman Coulter, citing A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or me….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1944-2020
ClassificationClass II
Statusterminated
Initiated2020-03-30
Published2020-05-20
Last updated2026-09-13 11:55 UTC
CompanyBeckman Coulter
DistributionUS

Product

iChem¿VELOCITY Urine Chemistry System-

Reason

A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.

Identifiers

code_infoCatalog #700-7176-001 and #700-7177-001, All part numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1944-2020%22