FDA Device recall Z-1943-2024

Arial mobile application, model numbers 54640, 54630, and 54630-SMA, that utilizes Apple iOS 17.4 and Apple iOS 17.5, a component of the Ar…

On 2024-06-12 the FDA classified a Class II recall of Arial mobile application, model numbers 54640, 54630, and 54630-SMA, that utilizes Apple iOS 17.4 and Apple iOS 17.5, a component of the Ar… by Securitas Healthcare, citing the Arial Mobile Application utilizing Apple iOS 17.4 and iOS 17.5 is frequently restarting causing the app to receive alerts for all pending alarm notifications and other events….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1943-2024
ClassificationClass II
Statusongoing
Initiated2024-04-11
Published2024-06-12
Last updated2026-09-13 11:57 UTC
CompanySecuritas Healthcare
DistributionUS

Product

Arial mobile application, model numbers 54640, 54630, and 54630-SMA, that utilizes Apple iOS 17.4 and Apple iOS 17.5, a component of the Arial Emergency and Nurse Call System.

Reason

The Arial Mobile Application utilizing Apple iOS 17.4 and iOS 17.5 is frequently restarting causing the app to receive alerts for all pending alarm notifications and other events when the app restarts, experience difficulty in placing and maintaining staff-to-staff phone calls, and frequently appears offline while the app restarts.

Identifiers

code_infoArial Mobile v11.0.0 UDI: (01)00840948100963(10)1100.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1943-2024%22