FDA Device recall Z-1941-2014
illumipro-10 Incubator/Reader, Meridian Bioscience, Inc
- FDA Device
- Class II
- terminated
- Published 2014-07-09
On 2014-07-09 the FDA classified a Class II recall of illumipro-10 Incubator/Reader, Meridian Bioscience, Inc by Meridian Bioscience, citing illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1941-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-04-22 |
| Published | 2014-07-09 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Meridian Bioscience |
| Distribution | AL, AZ, CA, CO, FL, IA, IN, LA, MA, ME, MI, MN, NJ, NY, OH, PA, WV |
Product
illumipro-10 Incubator/Reader, Meridian Bioscience, Inc. The illumipro-10 incubator/reader is designed to be used with illumigene Molecular Assays. The device incubates illumigene Test Devices and reads endpoint reactions.
Reason
Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration.
Identifiers
| code_info | Serial Numbers: IP2669, IP2670, IP2671, IP2672, IP2673, IP2674, IP2675, IP2676, IP2677, IP2678, IP2680, IP2682, IP2683…Serial Numbers: IP2669, IP2670, IP2671, IP2672, IP2673, IP2674, IP2675, IP2676, IP2677, IP2678, IP2680, IP2682, IP2683, IP2684, IP2685, IP2686, IP2687, IP2688, IP2689, IP2690, IP2691, IP2692, IP2693, IP2694, IP2695, IP2696, IP2697, IP2698, IP2699, IP2700, IP2701, IP2702, IP2703, IP2704, IP2705, IP2706. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Meridian Bioscience
- Revogene C. difficile, in vitro diagnostic assay intended for use with the Revogene instrument, Catalog Number 410300
FDA Device · Class II · 2026-01-14 · Meridian Bioscience - Meridian Bioscience, Group B Streptococcus, Catalog Number 480050, Nucleic Acid Amplification Test Assay
FDA Device · Class II · 2025-11-26 · Meridian Bioscience - Meridian Bioscience, alethia c. Difficile, Catalog Number 480050, Nucleic Acid Amplification Test Assay
FDA Device · Class II · 2025-11-26 · Meridian Bioscience - Para-Pak Zn-PVA/10% Formalin-for the routine collection, transportation, preservation, and examination of stool specimens for intestinal pa…
FDA Device · Class II · 2023-12-20 · Meridian Bioscience - Para-Pak Clean Vial-For the collection, transportation, preservation, and examination of stool specimens containing intestinal parasites Ca…
FDA Device · Class II · 2023-12-20 · Meridian Bioscience
All Meridian Bioscience recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1941-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.