RecallRadar

FDA Device recall Z-1940-2025

Femoral Nail, RIGHT T2 GTN 8x300 mm

On 2025-06-18 the FDA classified a Class II recall of Femoral Nail, RIGHT T2 GTN 8x300 mm by Stryker, citing investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version b….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1940-2025
ClassificationClass II
Statusongoing
Initiated2025-05-22
Published2025-06-18
Last updated2026-09-13 11:57 UTC
CompanyStryker
DistributionUS

Product

Femoral Nail, RIGHT T2 GTN 8x300 mm. Catalog Number: 1851-0830S.

Reason

Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.

Identifiers

code_infoCatalog Number: 1851-0830S. GTIN: 04546540647184. Batch lot number: K09F17F

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.