FDA Device recall Z-1940-2020
Formula 418 Renal Balloon-Expandable Stent, RPN FOR418-18-80-6-12, Order Number (GPN) G56611 Balloon-expandable stent made of 316L stainles…
- FDA Device
- Class II
- terminated
- Published 2020-05-20
On 2020-05-20 the FDA classified a Class II recall of Formula 418 Renal Balloon-Expandable Stent, RPN FOR418-18-80-6-12, Order Number (GPN) G56611 Balloon-expandable stent made of 316L stainles… by Cook, citing the firm received complaints on five devices from one specific lot where the stent dislodged from the balloon delivery system prior to deployment. Use of affected product may lead….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1940-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-04-15 |
| Published | 2020-05-20 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Cook |
| Distribution | IA, NC, OH, VA |
Product
Formula 418 Renal Balloon-Expandable Stent, RPN FOR418-18-80-6-12, Order Number (GPN) G56611 Balloon-expandable stent made of 316L stainless steel with a slotted tube configuration. It is pre-mounted on an over-the-wire balloon catheter, which serves as the delivery system. The stent is positioned between two radiopaque marker bands, which are located inside the balloon at the proximal and distal tapers of the balloon.
Reason
The firm received complaints on five devices from one specific lot where the stent dislodged from the balloon delivery system prior to deployment. Use of affected product may lead to increased procedural time and additional intervention to remove a stent that dislodged within an unintended location.
Identifiers
| code_info | Lot number 10153876 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1940-2020%22