RecallRadar

FDA Device recall Z-1934-2012

S4 Cervical (S4C) Revision Instruments (FW099R and FW101R). The S4C Revision Instruments assist in the removal of an occipital cervical scr…

On 2012-07-11 the FDA classified a Class II recall of S4 Cervical (S4C) Revision Instruments (FW099R and FW101R). The S4C Revision Instruments assist in the removal of an occipital cervical scr… by Aesculap Implant Systems, citing aesculap Implant Systems, LLC has initiated a voluntary removal of the S4 Cervical (S4C) Occiput Revision Instrument that is part of the S4C Occipital Instrument Set. The instrume….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1934-2012
ClassificationClass II
Statusterminated
Initiated2012-03-07
Published2012-07-11
Last updated2026-09-13 11:53 UTC
CompanyAesculap Implant Systems
DistributionUS

Product

S4 Cervical (S4C) Revision Instruments (FW099R and FW101R). The S4C Revision Instruments assist in the removal of an occipital cervical screw during revision surgery.

Reason

Aesculap Implant Systems, LLC has initiated a voluntary removal of the S4 Cervical (S4C) Occiput Revision Instrument that is part of the S4C Occipital Instrument Set. The instruments may not thread into the screw properly during a revision surgery due to tolerance issues between the two instruments.

Identifiers

code_infoAll distributed lot numbers and serial numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1934-2012%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.