RecallRadar

FDA Device recall Z-1931-2015

Uric Acid plus ver.2 In vitro test for the quantitative determination of uric acid in human serum, plasma and urine or Roche/Hitachi MODULA…

On 2015-07-08 the FDA classified a Class II recall of Uric Acid plus ver.2 In vitro test for the quantitative determination of uric acid in human serum, plasma and urine or Roche/Hitachi MODULA… by Roche Diagnostics Operations, citing possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1931-2015
ClassificationClass II
Statusterminated
Initiated2015-05-07
Published2015-07-08
Last updated2026-09-13 11:53 UTC
CompanyRoche Diagnostics Operations
DistributionUS

Product

Uric Acid plus ver.2 In vitro test for the quantitative determination of uric acid in human serum, plasma and urine or Roche/Hitachi MODULAR systems, cobas c111, COBAS Integra 800 and 400 plus systems, and cobas c systems.

Reason

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Identifiers

code_infoCatalog Numbers:04657608190; 03183807190; 05171857190; 03183807190; 11875426216; 11929429216; 11929437216.