FDA Device recall Z-1927-2015
LDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative determination of low-density li…
- FDA Device
- Class II
- terminated
- Published 2015-07-08
On 2015-07-08 the FDA classified a Class II recall of LDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative determination of low-density li… by Roche Diagnostics Operations, citing possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1927-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-05-07 |
| Published | 2015-07-08 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Roche Diagnostics Operations |
| Distribution | US |
Product
LDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative determination of low-density lipoprotein Cholesterol (LDL-C) in serum and plasma.
Reason
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
Identifiers
| code_info | Catalog numbers 04714423190; 04711220190; 05230438190; 05230446190. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1927-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.