FDA Device recall Z-1926-2018
Prismaflex 7.20 US, Product Code 955542 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (…
- FDA Device
- Class II
- terminated
- Published 2018-05-30
On 2018-05-30 the FDA classified a Class II recall of Prismaflex 7.20 US, Product Code 955542 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (… by Baxter Healthcare, citing baxter Healthcare will be installing new firmware on all Prismaflex control units to address the potential for a small number of these units to exhibit a failure mode with the pum….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1926-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-04-24 |
| Published | 2018-05-30 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Baxter Healthcare |
| Distribution | PR |
Product
Prismaflex 7.20 US, Product Code 955542 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. " Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated. " Continuous Renal Replacement Therapy (CRRT) in conjunction with the MARS system to conduct MARS treatments for patients weighing 20 kilograms or more. All treatments administered via the Prismaflex control unit must be prescribed by a physician.
Reason
Baxter Healthcare will be installing new firmware on all Prismaflex control units to address the potential for a small number of these units to exhibit a failure mode with the pump module electronics. The failure mode may result in a Voltage Out of Range malfunction alarm, which causes the device to enter a safe state and become inoperable until it is serviced. Baxter will be releasing new firmware that will prevent the malfunction from occurring.
Identifiers
| code_info | All Lots and Serial Numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.