FDA Device recall Z-1925-2018

Prismaflex 7.XX US, Product Code 115269 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (…

On 2018-05-30 the FDA classified a Class II recall of Prismaflex 7.XX US, Product Code 115269 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (… by Baxter Healthcare, citing baxter Healthcare will be installing new firmware on all Prismaflex control units to address the potential for a small number of these units to exhibit a failure mode with the pum….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1925-2018
ClassificationClass II
Statusterminated
Initiated2018-04-24
Published2018-05-30
Last updated2026-09-13 11:54 UTC
CompanyBaxter Healthcare
DistributionPR

Product

Prismaflex 7.XX US, Product Code 115269 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. " Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated. " Continuous Renal Replacement Therapy (CRRT) in conjunction with the MARS system to conduct MARS treatments for patients weighing 20 kilograms or more. All treatments administered via the Prismaflex control unit must be prescribed by a physician.

Reason

Baxter Healthcare will be installing new firmware on all Prismaflex control units to address the potential for a small number of these units to exhibit a failure mode with the pump module electronics. The failure mode may result in a Voltage Out of Range malfunction alarm, which causes the device to enter a safe state and become inoperable until it is serviced. Baxter will be releasing new firmware that will prevent the malfunction from occurring.

Identifiers

code_infoAll Lots and Serial Numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1925-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.