FDA Device recall Z-1925-2013

Dimension(R) Mass Creatine Kinase MB Isozyme Calibrator (RC420) The Mass Creatine Kinase MB isozyme Calibrator is an in vitro diagnostic pr…

On 2013-08-14 the FDA classified a Class II recall of Dimension(R) Mass Creatine Kinase MB Isozyme Calibrator (RC420) The Mass Creatine Kinase MB isozyme Calibrator is an in vitro diagnostic pr… by Siemens Healthcare Diagnostics, citing siemens Healthcare Diagnostics has received complaints that the lyophilized cake is not fully dissolved after the stated time in the MMB calibrator Instructions For Use (IFU). The….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1925-2013
ClassificationClass II
Statusterminated
Initiated2013-06-05
Published2013-08-14
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Dimension(R) Mass Creatine Kinase MB Isozyme Calibrator (RC420) The Mass Creatine Kinase MB isozyme Calibrator is an in vitro diagnostic product intended to be used to calibrate the Creatine Kinase MB Isozyme method for the Dimension(R) clinical chemistry system with heterogeneous immunoassay module.

Reason

Siemens Healthcare Diagnostics has received complaints that the lyophilized cake is not fully dissolved after the stated time in the MMB calibrator Instructions For Use (IFU). The frequency of this occurrence is low but if it occurs, gel-like clumps may be observed.

Identifiers

code_infoProduct - MMB CAL, Catalog Number RC420, Siemens Material Number 10445023 with lot numbers 2GD053, 2KD024, and 3AD051.