On 2017-05-10 the FDA classified a Class II recall of Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001 by Merge Healthcare, citing the units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1923-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-03-30 |
| Published | 2017-05-10 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Merge Healthcare |
| Distribution | AL, FL, ID, LA, MO, NC, OK, SC, TX, UT |
Product
Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system.
Reason
The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias added to the CO2 measurement and affect the alarm notification.
Identifiers
| code_info | ISA CO2 Sensor serial numbers: 107349, 107357, 107358, 107363, 107370, 107377, 107390, 107407, 107435, 107476, 107…ISA CO2 Sensor serial numbers: 107349, 107357, 107358, 107363, 107370, 107377, 107390, 107407, 107435, 107476, 107509, 107510, 107556, 107659, 107700, 107707, 107768, 107776, 107831, 107832, 107836, 107839, and 107841. |
|---|