On 2013-08-14 the FDA classified a Class II recall of bioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system by Biomerieux, citing A product issue has been identified for customers using software version 6.01 and are interfaced with a Laboratory Information System (LIS). During the update of software from ver….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1920-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-06-14 |
| Published | 2013-08-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Biomerieux |
| Distribution | US |
Product
bioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, UPDT V2S 6.01 ZH. Part number: 414569. The product is used with the VITEK 2 automated system which consists of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.
Reason
A product issue has been identified for customers using software version 6.01 and are interfaced with a Laboratory Information System (LIS). During the update of software from version 5.xx to 6.01, the Bid-directional Computer interface (BCI) antibiotic codes are not updated properly for Cefovecin, Prulifloxacin and Ceftaroline. For the referenced antibiotics, the firm code in the reference table
Identifiers
| code_info | version 6.01 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1920-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.