RecallRadar

FDA Device recall Z-1918-2020

REF: Z2850, 112" (284 cm) 10 Drop Primary set w/2 Clave, Remv 3 Gang 1o2 Manifold w/3 Clave, Baseplate, Rotating Luer w/Filter Cap, 1 Ext,…

On 2020-05-13 the FDA classified a Class II recall of REF: Z2850, 112" (284 cm) 10 Drop Primary set w/2 Clave, Remv 3 Gang 1o2 Manifold w/3 Clave, Baseplate, Rotating Luer w/Filter Cap, 1 Ext,… by Icu Medical, citing there is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrec….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1918-2020
ClassificationClass II
Statusterminated
Initiated2019-10-04
Published2020-05-13
Last updated2026-09-13 11:55 UTC
CompanyIcu Medical
DistributionAK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WY

Product

REF: Z2850, 112" (284 cm) 10 Drop Primary set w/2 Clave, Remv 3 Gang 1o2 Manifold w/3 Clave, Baseplate, Rotating Luer w/Filter Cap, 1 Ext, Sterile R, UDI: (01)00887709028241

Reason

There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.

Identifiers

code_infoLot number: 4172062

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1918-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.