REF: B9900-287, 82" (208 cm) Appx 10.4 ml, 10 Drop Admin Set, w/Pre-Slit Port, Clave, 1 Ext, Rotating Luer, Sterile R, UDI: (01)00887709016…

On 2020-05-13 the FDA classified a Class II recall of REF: B9900-287, 82" (208 cm) Appx 10.4 ml, 10 Drop Admin Set, w/Pre-Slit Port, Clave, 1 Ext, Rotating Luer, Sterile R, UDI: (01)00887709016… by Icu Medical, citing there is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrec….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1914-2020
ClassificationClass II
Statusterminated
Initiated2019-10-04
Published2020-05-13
Last updated2026-09-13 11:55 UTC
CompanyIcu Medical
DistributionAK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WY

Product

REF: B9900-287, 82" (208 cm) Appx 10.4 ml, 10 Drop Admin Set, w/Pre-Slit Port, Clave, 1 Ext, Rotating Luer, Sterile R, UDI: (01)00887709016842

Reason

There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.

Identifiers

code_infoLot number: 4167007

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1914-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.