FDA Device recall Z-1914-2016
VITROS 5600 Integrated System, Catalog Number 6802413, Unique Device Identifier No. 10758750002740
- FDA Device
- Class II
- terminated
- Published 2016-06-15
On 2016-06-15 the FDA classified a Class II recall of VITROS 5600 Integrated System, Catalog Number 6802413, Unique Device Identifier No. 10758750002740 by Ortho Clinical Diagnostics, citing ortho Clinical Diagnostics confirmed a software timing anomaly in which the VITROS 3600 and 5600 Systems may unexpectedly aspirate from or dispense into an unintended sample that….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1914-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-04-06 |
| Published | 2016-06-15 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Ortho Clinical Diagnostics |
| Distribution | US |
Product
VITROS 5600 Integrated System, Catalog Number 6802413, Unique Device Identifier No. 10758750002740; and VITROS 5600 Integrated System (refurbished), Catalog Number 6802915, Unique Device Identifier No. 10758750007110; IVD.
Reason
Ortho Clinical Diagnostics confirmed a software timing anomaly in which the VITROS 3600 and 5600 Systems may unexpectedly aspirate from or dispense into an unintended sample that may lead to erroneous or contaminated results.
Identifiers
| code_info | Software Version 3.2.2 & below; VITROS 5600 Serial Numbers J56000024 - J56002387; VITROS 5600 (Refurbished) US Serial N…Software Version 3.2.2 & below; VITROS 5600 Serial Numbers J56000024 - J56002387; VITROS 5600 (Refurbished) US Serial Numbers: 56000186, 56000308, 56000315, 56000327, 56000339, 56000458, 56000559, 56000718, 56000727, 56000741, 56000747, 56000750, 56000786, 56000792, 56000824, 56001062, 56001235, 56001635, 56001641; Foreign Serial Numbers: 56000147, 56000200, 56000214, 56000245,56000283, 56000330, 56000331, 56000362, 56000377, 56000399, 56000413, 56000416, 56000428, 56000437, 56000438, 56000439, 56000459, 56000463, 56000464, 56000466, 56000515, 56000528, 56000537, 56000555, 56000560, 56000632, 56000642, 56000669, 56000680, 56000749, 56000867, 56001133, 56001135, 56001246, 56001461, 56001482, 56001487, 56001859. Unique Device Identifier No. 10758750002740 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1914-2016%22