DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument

On 2016-06-15 the FDA classified a Class II recall of DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument by Synthes Usa Products, citing devices did not pass the biological safety evaluation for cytotoxicity following exposure to test conditions. The high growth inhibition levels observed during testing could be at….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1912-2016
ClassificationClass II
Statusterminated
Initiated2016-05-02
Published2016-06-15
Last updated2026-09-13 11:54 UTC
CompanySynthes Usa Products
DistributionUS

Product

DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument. Part number 351.43

Reason

Devices did not pass the biological safety evaluation for cytotoxicity following exposure to test conditions. The high growth inhibition levels observed during testing could be attributed to corrosion of the device at solder points. This could potentially be reproduced during use and reprocessing.

Identifiers

code_info
Catalog ID # 351.43 Lot numbers: XXX5071, 3945479, 4738380, XXX5072, 3997000, 4816738, XXX5073, 3996999, 4816739, A3…Catalog ID # 351.43 Lot numbers: XXX5071, 3945479, 4738380, XXX5072, 3997000, 4816738, XXX5073, 3996999, 4816739, A3CN449, 3997338, 4441056, 3007721, 4326050, 5407442, 3148892, 4307973, 5582819, 3175916, 4387338, 5722664, 3175202, 4441055, 5813837, 3175201, 4441079, 5786778, 3921562, 4559672, 5800844, 3928332, 4561383, 5903413, 3935789, 4619549, 5910995.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1912-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.