RecallRadar

FDA Device recall Z-1900-2019

Siemens Artis Q biplane---Interventional Fluoroscopic X-Ray System Model Number: 10848282

On 2019-07-10 the FDA classified a Class II recall of Siemens Artis Q biplane---Interventional Fluoroscopic X-Ray System Model Number: 10848282 by Siemens Medical Solutions Usa, citing artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1900-2019
ClassificationClass II
Statusterminated
Initiated2019-06-06
Published2019-07-10
Last updated2026-09-13 11:55 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

Siemens Artis Q biplane---Interventional Fluoroscopic X-Ray System Model Number: 10848282

Reason

Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel

Identifiers

code_info
Serial Numbers: 121013 121057 121056 121058 121006 121185 121016 121027 121020 121321 121448 121040 121049…Serial Numbers: 121013 121057 121056 121058 121006 121185 121016 121027 121020 121321 121448 121040 121049 121015 121094 121082 121089 121383 121144 121319 121137 121085 121155 121193 121136 121165 121140 121141 121151 121168 121169 121154 121211 121188 121415 121183 121209 121187 121309 121200 121313 121206 121361 121366 121359 121393 121301 121407 121406 121352 121379 121364 121342 121471 121520 121411 121412 121402 121430 121502 121421 121472 121460 121457 121515 121444 121516 121510 121491 121525 121530 121527

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1900-2019%22