FDA Device recall Z-1899-2017
Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android Product Usage: This is a digital product avai…
- FDA Device
- Class II
- terminated
- Published 2017-05-10
On 2017-05-10 the FDA classified a Class II recall of Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android Product Usage: This is a digital product avai… by Roche Diabetes Care, citing A program error (bug) in the Bolus Advisor feature, which could result in incorrect bolus advice and a potential insulin over-delivery.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1899-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-03-14 |
| Published | 2017-05-10 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Roche Diabetes Care |
| Distribution | US |
Product
Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android Product Usage: This is a digital product available for download direct to customers from the Apple App Store and the Google Play Store. The total number of downloads is not available. The number of unique users for each affected version from the date of initial distribution through 03/23/2017 is noted in 7a. However, some users will have used all three affected versions as they upgraded from version to version when the versions became available and many others will have tried the app and discontinued use. In the last 30 days, 02/22/2017 03/23/2017, there have been 16,781 unique users in the US with 264 of those seeking bolus advice and 19,268 unique users in the rest of world (excluding Canada) with 415 of those users seeking bolus advice.
Reason
A program error (bug) in the Bolus Advisor feature, which could result in incorrect bolus advice and a potential insulin over-delivery.
Identifiers
| code_info | Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1899-2017%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.