FDA Device recall Z-1896-2021
OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine
- FDA Device
- Class II
- terminated
- Published 2021-06-23
On 2021-06-23 the FDA classified a Class II recall of OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine by Ossur Americas, citing due to insufficient column strength of the racks for all load conditions, racks associated with height adjustment functionality have been observed deforming, which may lead to inc….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1896-2021 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-06-01 |
| Published | 2021-06-23 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Ossur Americas |
| Distribution | US |
Product
OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine. REF: MJS-101; MJSR-101
Reason
Due to insufficient column strength of the racks for all load conditions, racks associated with height adjustment functionality have been observed deforming, which may lead to increased range of motion in the cervical spine.
Identifiers
| code_info | REF Number: MJS-101; MJSR-101. Lot Numbers: MX180425 through MX200306 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1896-2021%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.