RecallRadar

FDA Device recall Z-1896-2021

OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine

On 2021-06-23 the FDA classified a Class II recall of OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine by Ossur Americas, citing due to insufficient column strength of the racks for all load conditions, racks associated with height adjustment functionality have been observed deforming, which may lead to inc….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1896-2021
ClassificationClass II
Statusterminated
Initiated2021-06-01
Published2021-06-23
Last updated2026-09-13 11:56 UTC
CompanyOssur Americas
DistributionUS

Product

OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine. REF: MJS-101; MJSR-101

Reason

Due to insufficient column strength of the racks for all load conditions, racks associated with height adjustment functionality have been observed deforming, which may lead to increased range of motion in the cervical spine.

Identifiers

code_infoREF Number: MJS-101; MJSR-101. Lot Numbers: MX180425 through MX200306

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1896-2021%22