FDA Device recall Z-1894-2025
Signature Laparoscopic Instruments, Allis Clamp Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector
- FDA Device
- Class II
- ongoing
- Published 2025-06-11
On 2025-06-11 the FDA classified a Class II recall of Signature Laparoscopic Instruments, Allis Clamp Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector by Silex Medical, citing insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1894-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-01-09 |
| Published | 2025-06-11 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Silex Medical |
| Distribution | US |
Product
Signature Laparoscopic Instruments, Allis Clamp Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5862030.
Reason
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
Identifiers
| code_info | Model/REF Number: 5862030. UDI-DI: B33158620300. Lot Number: M28182. Serial Numbers: 0709, 0710, 0711, 0712, 0713, 0714, 0715, 0716, 0717, 0718, 0719, 0720 |
|---|