RecallRadar

FDA Device recall Z-1892-2025

BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound tra…

On 2025-06-11 the FDA classified a Class II recall of BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound tra… by Civco Medical Instruments, citing during the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1892-2025
ClassificationClass II
Statusongoing
Initiated2025-05-01
Published2025-06-11
Last updated2026-09-13 11:57 UTC
CompanyCivco Medical Instruments
DistributionGA

Product

BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers

Reason

During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.

Identifiers

code_info
UDI-DI: 00841436120050; Lot Numbers: A174424, A176855, A177208, A182744, A184023, A185336, A185337, A186904, A189734, A…UDI-DI: 00841436120050; Lot Numbers: A174424, A176855, A177208, A182744, A184023, A185336, A185337, A186904, A189734, A190489, A193623, A194755, A198716, A198795, A199940, A200430, A202208, A203169, A203201, A203520, A205457, A207901, A208740, A211619, A212291, A213027, A215531, A216885, A217412, A220102, A221365, A221925, A225145, A226800, A228446, A228778, A229092, A229898, A230045, A230054, A230549, A230581, A231101, A231810, A232903, A233075, A233535, A236538, A238767, A239339, A239355, A240873, A240931, A241445, A242018, A242161, A243263, A243264, A245342, A245527, A246173, A248520, A248572, A250325, A252513, A252842, A254080, A254081, A254900, A254955, A255332, A257534, A258150, A258926, A260607

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1892-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.