RecallRadar

FDA Device recall Z-1890-2024

Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320

On 2024-06-05 the FDA classified a Class II recall of Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320 by Boston Scientific Neuromodulation, citing deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1890-2024
ClassificationClass II
Statusongoing
Initiated2024-04-18
Published2024-06-05
Last updated2026-09-13 11:57 UTC
CompanyBoston Scientific Neuromodulation
DistributionUS

Product

Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320

Reason

Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.

Identifiers

code_info
REF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365DB12160/8714729985044/100209 - 753347/09-OCT-2020 throu…REF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365DB12160/8714729985044/100209 - 753347/09-OCT-2020 through 26-MAR-2026; M365DB12320/8714729985051/100104 - 753200/09-OCT-2020 through 21-MAR-2026

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1890-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.