FDA Device recall Z-1889-2020
Siemens Aptio Automation Centrifuge Module, tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-…
- FDA Device
- Class II
- ongoing
- Published 2020-05-13
On 2020-05-13 the FDA classified a Class II recall of Siemens Aptio Automation Centrifuge Module, tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-… by Inpeco, citing the FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1889-2020 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2020-04-07 |
| Published | 2020-05-13 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Inpeco |
| Distribution | AL, CA, CT, DE, FL, IA, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NY, OH, PA, RI, TN, TX, VA, WA |
Product
Siemens Aptio Automation Centrifuge Module, tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf
Reason
The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Identifiers
| code_info | Serial Numbers: AP2.0001, AP2.0002, AP2.0005, AP2.0008, AP2.0010, AP2.0011, AP2.0012, AP2.0013, AP2.0014, AP2…Serial Numbers: AP2.0001, AP2.0002, AP2.0005, AP2.0008, AP2.0010, AP2.0011, AP2.0012, AP2.0013, AP2.0014, AP2.0016, AP2.0018, AP2.0019, AP2.0023, AP2.0024, AP2.0025, AP2.0026, AP2.0027, AP2.0032, AP2.0033, AP2.0034, AP2.0036, AP2.0039, AP2.0041, AP2.0042, AP2.0044, AP2.0045, AP2.0046, AP2.0047, AP2.0049, AP2.0050, AP2.0051, AP2.0052, AP2.0053, AP2.0054, AP2.0055, AP2.0056, AP2.0057, AP2.0058, AP2.0059, AP2.0060, AP2.0061, AP2.0062, AP2.0063, AP2.0064, AP2.0065, AP2.0067, AP2.0068, AP2.0070, AP2.0071, AP2.0072, AP2.0073, AP2.0074, AP2.0075, AP2.0076, AP2.0077, AP2.0078, AP2.0080, AP2.0081, AP2.0082, AP2.0083, AP2.0084, AP2.0085, AP2.0087, AP2.0088, AP2.0089, AP2.0090, AP2.0091, AP2.0092, AP2.0093, AP2.0094, AP2.0095, AP2.0097, AP2.0098, AP2.0099, AP2.0100, AP2.0101, AP2.0103, AP2.0105, AP2.0106, AP2.0107, AP2.0108, AP2.0109, AP2.0110, AP2.0111, AP2.0112, AP2.0113, AP2.0114, AP2.0115, AP2.0116, AP2. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1889-2020%22