On 2012-07-04 the FDA classified a Class II recall of CPB CATHETER KIT-DR GERHARDT Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in t… by Terumo Cardiovascular Systems, citing during production of the Sarns Antegrade Cardioplegia Cannula, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of a foreign substance on the inner surface of so….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1889-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-05-09 |
| Published | 2012-07-04 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | US |
Product
CPB CATHETER KIT-DR GERHARDT Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during cardiopulmonary bypass procedures.
Reason
During production of the Sarns Antegrade Cardioplegia Cannula, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of a foreign substance on the inner surface of some cannulae tips. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface. The substance was likely deposited during the molding process, but the exact composition of the
Identifiers
| code_info | Catalog number: 80001-02 and lot number: NP20 and PA16. |
|---|