FDA Device recall Z-1887-2019
EPi-Sense Guided Coagulation System with Visitrax, MOdel No. CDK-1413 - Product Usage: The EPi-Sense Guided Coagulation System with VisiTra…
- FDA Device
- Class II
- terminated
- Published 2019-07-03
On 2019-07-03 the FDA classified a Class II recall of EPi-Sense Guided Coagulation System with Visitrax, MOdel No. CDK-1413 - Product Usage: The EPi-Sense Guided Coagulation System with VisiTra… by Atricure, citing the sterile package seal for the Epi-Sense Guided Coagulation System may be compromised, and has a worst-case reasonable harm of infection if used.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1887-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-05-29 |
| Published | 2019-07-03 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Atricure |
| Distribution | US |
Product
EPi-Sense Guided Coagulation System with Visitrax, MOdel No. CDK-1413 - Product Usage: The EPi-Sense Guided Coagulation System with VisiTrax, consists of a sterile, single-use, disposable coagulation electrode device (1cm, 2cm, and 3cm sizes provided) intended to be used to coagulate cardiac tissue. The flexible, cooled electrode device, with a suction stabilizer feature, transmits radiofrequency (RF) energy from an Electrosurgical Generator (non-sterile, re-useable) connected through an Instrument Cable (sterile). A temporary sensing electrode feature may be used with an off-the-shelf electrogram recording system with the use of an additional instrument cable (non-sterile). An accessory Cannula may be used to facilitate coagulation device access and visibility of the heart. The EPi-Sense Guided Coagulation System with VisiTrax technology is intended for the coagulation of cardiac tissue using Radiofrequency (RF) energy using thoracoscopic, endoscopic, and laparoscopic surgical techniques and may be used for temporary cardiac signal sensing and recording during surgery when connected to a temporary external recording device.
Reason
The sterile package seal for the Epi-Sense Guided Coagulation System may be compromised, and has a worst-case reasonable harm of infection if used.
Identifiers
| code_info | Lots 70638, 71332, 89208, 89938, 90624 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1887-2019%22