RecallRadar

FDA Device recall Z-1878-2016

Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44

On 2016-06-08 the FDA classified a Class II recall of Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44 by Natus Medical, citing the replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic gas when heated, and can drip down durin….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1878-2016
ClassificationClass II
Statusterminated
Initiated2016-04-22
Published2016-06-08
Last updated2026-09-13 11:54 UTC
CompanyNatus Medical
DistributionAZ, CA, DC, FL, IL, KS, MA, MD, MN, NC, NH, NY, OH, OK, PA, VA, WA, WI

Product

Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue Part Number: 54343, 54344, and 100105-01.

Reason

The replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic gas when heated, and can drip down during use, contaminating the interior chamber of the Sterile Drier and its contents.

Identifiers

code_info
Filter Serial Numbers: Filter SN: 2522893, 2522897, 2522898, 2522899, 2522900, 2522903, 2522914, 2522915, 2522916, 2522…Filter Serial Numbers: Filter SN: 2522893, 2522897, 2522898, 2522899, 2522900, 2522903, 2522914, 2522915, 2522916, 2522917, 2522918, 2522919, 2522920, 2522921, 2522922, 2522923, 2522924, 2522925, 2522926, 2522930, and 2522931.