RecallRadar

FDA Device recall Z-1876-2019

ECG leadwire sets: REF/Catalog Number Description 2106389-001 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106389-002 ECG…

On 2019-07-03 the FDA classified a Class III recall of ECG leadwire sets: REF/Catalog Number Description 2106389-001 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106389-002 ECG… by Ge Healthcare, citing ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1876-2019
ClassificationClass III
Statusterminated
Initiated2019-05-10
Published2019-07-03
Last updated2026-09-13 11:55 UTC
CompanyGe Healthcare
DistributionUS

Product

ECG leadwire sets: REF/Catalog Number Description 2106389-001 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106389-002 ECG Leadwire set, 5-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106389-003 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106389-004 ECG Leadwire set, 5-lead, grouped, grabber, IEC, 130 cm/ 51 in 2106389-005 ECG Leadwire set, 5-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106389-006 ECG Leadwire set, 5-lead, grouped, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106390-001 ECG Leadwire set, 3-lead, grabber, AHA, 74 cm/ 29 in 2106390-002 ECG Leadwire set, 3-lead, grabber, AHA, 130 cm/ 51 in 2106390-003 ECG Leadwire set, 3-lead, grabber, IEC, 74 cm/ 29 in 2106390-004 ECG Leadwire set, 3-lead, grabber, IEC, 130 cm/ 51 in 2106391-001 ECG Leadwire set, 5-lead, grabber, AHA, 74 cm/ 29 in 2106391-002 ECG Leadwire set, 5-lead, grabber, AHA, 130 cm/ 51 in 2106391-003 ECG Leadwire set, 5-lead, grabber, IEC, 74 cm/ 29 in 2106391-004 ECG Leadwire set, 5-lead, grabber, IEC, 130 cm/ 51 in 2106391-005 ECG Leadwire set, 5-lead, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106391-006 ECG Leadwire set, 5-lead, grabber, IEC, mix 74 cm/ 29 in, 130 cm/ 51 in 2106393-001 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 130 cm/ 51 in 2106393-002 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 130 cm/ 51 in 2106393-003 ECG Leadwire set, 5-lead V2-6, grabber, AHA, 74 cm/ 29 in 2106393-004 ECG Leadwire set, 5-lead C2-6, grabber, IEC, 74 cm/ 29 in 2106397-001 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 74 cm/ 29 in 2106397-002 ECG Leadwire set, 6-lead, grouped, grabber, AHA, 130 cm/ 51 in 2106397-003 ECG Leadwire set, 6-lead, grouped, grabber, AHA, mix 74 cm/ 29 in, 130 cm/ 51 in 2106398-001 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 74 cm/ 29 in 2106398-002 ECG Leadwire set, 6-lead, grouped, grabber, IEC, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

Reason

ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

Identifiers

code_info
product does not have lot or serial codes REF/Catalog Number GTIN/UDI 2106389-001 00840682139960 2106389-002 00840…product does not have lot or serial codes REF/Catalog Number GTIN/UDI 2106389-001 00840682139960 2106389-002 00840682139779 2106389-003 none 2106389-004 none 2106389-005 00840682139786 2106389-006 none 2106390-001 00840682139687 2106390-002 00840682140232 2106390-003 none 2106390-004 none 2106391-001 00840682140799 2106391-002 00840682139830 2106391-003 none 2106391-004 none 2106391-005 00840682140348 2106391-006 none 2106393-001 00840682140287 2106393-002 none 2106393-003 00840682139885 2106393-004 none 2106397-001 00840682140522 2106397-002 00840682140461 2106397-003 00840682140041 2106398-001 none 2106398-002 none

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1876-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.