RecallRadar

FDA Device recall Z-1871-2019

Thermo Scientific General Purpose Dynamic Bath Precision 5 & 10L Dual GP Bath Model Number: TSGP15D

On 2019-07-03 the FDA classified a Class II recall of Thermo Scientific General Purpose Dynamic Bath Precision 5 & 10L Dual GP Bath Model Number: TSGP15D by Thermo Fisher Scientific Asheville, citing product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1871-2019
ClassificationClass II
Statusterminated
Initiated2019-04-02
Published2019-07-03
Last updated2026-09-13 11:55 UTC
CompanyThermo Fisher Scientific Asheville
DistributionCA, GA, KY, PA

Product

Thermo Scientific General Purpose Dynamic Bath Precision 5 & 10L Dual GP Bath Model Number: TSGP15D

Reason

Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.

Identifiers

code_infoSerial Numbers: 300246816 300246817 300247411 300247412

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1871-2019%22