Thermo Scientific General Purpose Dynamic Bath Precision 10L GP Bath Model Number: TSGP10

On 2019-07-03 the FDA classified a Class II recall of Thermo Scientific General Purpose Dynamic Bath Precision 10L GP Bath Model Number: TSGP10 by Thermo Fisher Scientific Asheville, citing product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1868-2019
ClassificationClass II
Statusterminated
Initiated2019-04-02
Published2019-07-03
Last updated2026-09-13 11:55 UTC
CompanyThermo Fisher Scientific Asheville
DistributionCA, GA, KY, PA

Product

Thermo Scientific General Purpose Dynamic Bath Precision 10L GP Bath Model Number: TSGP10

Reason

Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.

Identifiers

code_infoSerial Numbers: 300246811 300246812 300246814 300246815 300247403 300247404 300247406 300247407 300247409 300247789 300247791 300247792