FDA Device recall Z-1866-2014
LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, fi…
- FDA Device
- Class II
- terminated
- Published 2014-07-02
On 2014-07-02 the FDA classified a Class II recall of LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, fi… by Lap, citing while projecting yellow lines it was notice that the yellow projection has the potential to splits into two separate red and green projections. In this case the difference between….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1866-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-03-21 |
| Published | 2014-07-02 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Lap |
| Distribution | NC |
Product
LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, field and MLC contours along with a 3D patient view.
Reason
While projecting yellow lines it was notice that the yellow projection has the potential to splits into two separate red and green projections. In this case the difference between the red and green projection was up to 2.5mm.
Identifiers
| code_info | Product Code: 0007569-0001 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1866-2014%22