FDA Device recall Z-1865-2019
Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 5L GP Bath Model Number: FSGPD05
- FDA Device
- Class II
- terminated
- Published 2019-07-03
On 2019-07-03 the FDA classified a Class II recall of Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 5L GP Bath Model Number: FSGPD05 by Thermo Fisher Scientific Asheville, citing product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1865-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-04-02 |
| Published | 2019-07-03 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Thermo Fisher Scientific Asheville |
| Distribution | CA, GA, KY, PA |
Product
Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 5L GP Bath Model Number: FSGPD05
Reason
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Identifiers
| code_info | Serial Numbers: 30024679S 300246796 300246797 300246798 300246799 300247393 300247394 30024739S 300247396 3002477S8 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1865-2019%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.