RecallRadar

FDA Device recall Z-1865-2017

Vacutainer Tubes Vacutainer tubes are intended for clinical laboratory hematology studies including but not limited to lead level testing,…

On 2017-04-26 the FDA classified a Class II recall of Vacutainer Tubes Vacutainer tubes are intended for clinical laboratory hematology studies including but not limited to lead level testing,… by Avid Medical, citing avid Medical Inc has received a complaint was received from an end-user indicating the vacutainer component 519700 (BD: 367863) expiring 12/16 was discovered inside of a kit label….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1865-2017
ClassificationClass II
Statusterminated
Initiated2017-03-15
Published2017-04-26
Last updated2026-09-13 11:54 UTC
CompanyAvid Medical
Distributionn/a

Product

Vacutainer Tubes Vacutainer tubes are intended for clinical laboratory hematology studies including but not limited to lead level testing, FEP level testing, and CBC testing.

Reason

Avid Medical Inc has received a complaint was received from an end-user indicating the vacutainer component 519700 (BD: 367863) expiring 12/16 was discovered inside of a kit labelled with an expiration date after 12/16. Upon receipt of complaint, AVID Medical promptly initiated an internal investigation and root cause analysis. Investigation found mixed component lots were contained in product 519700 component case. This occurrence resulted in the incorrect lot number and expiration being attributed to all work orders produced using the affected component. This issue was immediately contained and corrected. A full inventory analysis was conducted reviewing all vacutainer component inventories. The issue was found to affect a total of five vacutainer components.

Identifiers

code_info
Part Number/Product Number ILEY029-01 ULKY005-14 LOUS015-03 DNAZ017-01 ILEY029-02 SERA003-07 TENN016-05 DON…Part Number/Product Number ILEY029-01 ULKY005-14 LOUS015-03 DNAZ017-01 ILEY029-02 SERA003-07 TENN016-05 DONS008-05 EVER007 SERA005-07 LANC029-06 STLM007-10 CCHM024-03 Part Number/Product Number LANC015-27 CCHM027-04 STLM003-11 ABIN137-01 ABIN110-01 LANC015-28 MONT045-03 ILEY003-53 NAFB015-07 ILEY014-21 ILEY034-04 ILEY034-05 Part Number/Product Number AV10654-10 ILEY014-22 EVER010 EVER001 SOHS016-03 SSHH118-03 SSHH117-06 SOHS017-03 HCMH009-05 WIME023-08 TUFT040-05 Part Number/Product Number TUFT039-08 HCMH009-06 GIFT022-01 TUFT040-06 PCHM004-07 HLDH008-04 WHWC005-17 PCHM008-04 LANC015-25 PCHM011-03