RecallRadar

FDA Device recall Z-1865-2016

The ARCHITECT Estradiol Kit Intended to measure estradiol, an estrogenic steroid, in plasma

On 2016-06-08 the FDA classified a Class II recall of The ARCHITECT Estradiol Kit Intended to measure estradiol, an estrogenic steroid, in plasma by Abbott Laboratories, citing interaction of Fulvestrant with the ARCHITECT Estradiol assay, leading to falsely elevated estradiol results in patients treated with Fulvestrant.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1865-2016
ClassificationClass II
Statusterminated
Initiated2016-03-24
Published2016-06-08
Last updated2026-09-13 11:54 UTC
CompanyAbbott Laboratories
DistributionAK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY

Product

The ARCHITECT Estradiol Kit Intended to measure estradiol, an estrogenic steroid, in plasma.

Reason

Interaction of Fulvestrant with the ARCHITECT Estradiol assay, leading to falsely elevated estradiol results in patients treated with Fulvestrant.

Identifiers

code_info
Product List Number 07K7220 with Lot #: 55900UI01with expiration date of 04/23/2016, Lot #: 55908UI01 with expiration d…Product List Number 07K7220 with Lot #: 55900UI01with expiration date of 04/23/2016, Lot #: 55908UI01 with expiration date of 04/29/2016, Lot #: 55941UI00 with expiration date of 04/23/2016, Lot #: 57929UI00 with expiration date of 07/16/2016, Lot #: 57931UI01 with expiration date of 07/27/2016, Lot #: 60105UI00 with expiration date of 10/02/2016 and Lot #: 61106UI00 with expiration date of 11/27/2016. Product List Number 07K7225 - Lot #: 55900UI00 with expiration date of 04/23/2016, Lot #: 55908UI00 with expiration date of 04/29/2016, Lot #: 57929UI01 with expiration date of 07/16/2016, Lot #: 57931UI00 with expiration date of 08/27/2016, Lot #: 60105UI01 with expiration date of 02/10/2016, Lot #: 61106UI01 with expiration date of 11/27/2016 and Lot #: 62263UI00 with expiration date of 11/27/2016.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1865-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.