FDA Device recall Z-1855-2018
(1)BOX PST04089 SLEEVE AIWA CLINIC Item Number:PST04089 SLEEVE AIWA CLINIC (2)BOX PST04090 BYPASS AIWA CLINIC Item Number: PST04090 BYPASS…
- FDA Device
- Class II
- terminated
- Published 2018-05-23
On 2018-05-23 the FDA classified a Class II recall of (1)BOX PST04089 SLEEVE AIWA CLINIC Item Number:PST04089 SLEEVE AIWA CLINIC (2)BOX PST04090 BYPASS AIWA CLINIC Item Number: PST04090 BYPASS… by Covidien Medtronic, citing device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading t….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1855-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-04-26 |
| Published | 2018-05-23 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Covidien Medtronic |
| Distribution | US |
Product
(1)BOX PST04089 SLEEVE AIWA CLINIC Item Number:PST04089 SLEEVE AIWA CLINIC (2)BOX PST04090 BYPASS AIWA CLINIC Item Number: PST04090 BYPASS AIWA CLINIC Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simultaneously divides the tissue between the two, triple-staggered rows. The size of the staples is determined by the selection of the 2.0 mm, 2.5 mm, 3.5 mm or 4.8 mm single use loading unit. The Endo GIA universal staplers will accommodate any of the single use loading unit sizes that are available in the 30 mm, 45 mm and 60 mm lines.
Reason
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Identifiers
| code_info | Lot Numbers: (1) 0215237526, 0214991351, 0214991350, 0215157998, 0215157997 (2) 0215237527, 0215282382, 0214991349, 0214991348, 0214991347 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1855-2018%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.