FDA Device recall Z-1854-2024

MR 5300- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device

On 2024-05-29 the FDA classified a Class II recall of MR 5300- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device by Philips North America, citing the g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resultin….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1854-2024
ClassificationClass II
Statusongoing
Initiated2024-05-03
Published2024-05-29
Last updated2026-09-13 11:57 UTC
CompanyPhilips North America
DistributionUS

Product

MR 5300- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 782152 (2) 782110

Reason

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Identifiers

code_info
UDI-DI: (1) 782152 N/A (2) 782110 (01)00884838099364(21) Serial Numbers: 79113 79118 79127 79157 79122 79147 79105 7…UDI-DI: (1) 782152 N/A (2) 782110 (01)00884838099364(21) Serial Numbers: 79113 79118 79127 79157 79122 79147 79105 79150 79152 79183 79192 79102 79108 79112 79115 79116 79153 79156 79162 79163 79164 79168 79188 79189 79100 79126 79179 79143 79159 79104 79132 79160 79182 79165 79107 79158 79175 79136 79138 79139 79184 79176 79180 79171 79172 79111 79125 79134 79135 79148 79177 79181 79173 79187 79190 79133 79110 79101 79117 79166 79124 79123 79130 79174 79178 79151 79114 79103 79106 79121 79128 79129 79137 79140 79155 79161 79167 79169 79170 79185 Serial Number: 30199 30200 30201 30062 30191 30240 30241 30242 30305 30315 30125 30088 30113 30222 30252 30268 30028 30038 30058 30069 30081 30095 30100 30117 30118 30138 30164 30167 30168 30169 30170 30192 30197 30207 30209 30217 30228 30246 30248 30249 30256 30262 30267 30281 30284 30289 30297 30298 30306 30307 30312 30314 30210 30219 30245 30254 30276 30029 30043 30052 30061 30083 30101 30107 30114 30115 30116 30119 30120 30122 30130