FDA Device recall Z-1854-2019
Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568811911 Model: VLT600SF AIM STP
- FDA Device
- Class II
- terminated
- Published 2019-07-03
On 2019-07-03 the FDA classified a Class II recall of Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568811911 Model: VLT600SF AIM STP by Maquet Cardiovascular Us Sales, citing potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may r….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1854-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-05-28 |
| Published | 2019-07-03 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Maquet Cardiovascular Us Sales |
| Distribution | US |
Product
Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568811911 Model: VLT600SF AIM STP
Reason
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
Identifiers
| code_info | Serial Numbers: 31000 31001 31002 31003 31004 31006 31008 31009 31010 31012 31017 31018 31019 31020 31021 31022 3…Serial Numbers: 31000 31001 31002 31003 31004 31006 31008 31009 31010 31012 31017 31018 31019 31020 31021 31022 31023 31024 31025 31026 35000 35001 35011 35012 35013 35014 35015 35017 35018 35019 35020 35021 35022 35023 35024 35025 35026 35027 35028 35030 35031 35032 35033 38000 38001 38002 38003 38009 38011 38012 38013 38014 38015 38016 38018 38019 38020 38021 38022 38023 38025 38026 38027 |
|---|