FDA Device recall Z-1854-2014
NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-301 Lactate 45, Potassium 1, Sodium 140, Calcium 3, Magnesi…
- FDA Device
- Class II
- terminated
- Published 2014-07-02
On 2014-07-02 the FDA classified a Class II recall of NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-301 Lactate 45, Potassium 1, Sodium 140, Calcium 3, Magnesi… by Nxstage Medical, citing SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1854-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-05-14 |
| Published | 2014-07-02 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Nxstage Medical |
| Distribution | US |
Product
NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-301 Lactate 45, Potassium 1, Sodium 140, Calcium 3, Magnesium 1, Chloride 100, Glucose (mg/dL) 100 Batch Size (liters) 60 The SAK is one of the disposables used with NxStage PureFlow SL Dialysate Preparation System. The SAK is a disposable bag pre-filled with dialysate concentrate that is then filled with purified water prepared by NxStage PureFlow SL Dialysate Preparation System to prepare a batch of dialysate for use with the NxStage System One. SAKs are packaged in a case containing two SAKs. SAK product SKUs vary in the dialysate concentrate and volume of the final dialysate batch.
Reason
SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification
Identifiers
| code_info | Lot codes: 40279015 40179163 40179162 31279127 3067903 3057931 3057930 3047933 3047928 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1854-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.