RecallRadar

FDA Device recall Z-1848-2015

Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System with catalogue number SCF-MLD265 and SCF-MLD273

On 2015-07-08 the FDA classified a Class II recall of Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System with catalogue number SCF-MLD265 and SCF-MLD273 by Pega Medical, citing SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise ro….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1848-2015
ClassificationClass II
Statusterminated
Initiated2015-06-15
Published2015-07-08
Last updated2026-09-13 11:53 UTC
CompanyPega Medical
DistributionFL, TN

Product

Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System with catalogue number SCF-MLD265 and SCF-MLD273.

Reason

SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise rotation.

Identifiers

code_infoPart Number: SCF-MLD265, LOT#: 130703-01. Part Number: SCF-MLD273, LOT#: 130703-02.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.