FDA Device recall Z-1847-2023
Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504560
- FDA Device
- Class I
- ongoing
- Published 2023-07-05
On 2023-07-05 the FDA classified a Class I recall of Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504560 by Teleflex, citing reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1847-2023 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2023-05-25 |
| Published | 2023-07-05 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Teleflex |
| Distribution | US |
Product
Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504560
Reason
Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
Identifiers
| code_info | UDI/DI 4026704547593, Batch Numbers: 18FG03, 18FG15, 18FG28, 18GG24, 18IG27, 18JG02, 18JG17, 18JG26, 18JG34, 18KG09, 1…UDI/DI 4026704547593, Batch Numbers: 18FG03, 18FG15, 18FG28, 18GG24, 18IG27, 18JG02, 18JG17, 18JG26, 18JG34, 18KG09, 18KG35, 18LG05, 18LG12, 18LG18, 18LG37, 19AG27, 19AG37, 19BG12, 19CT32, 19CT41, 19CT55, 19CT58, 19ET66, 19ET77, 19FT14, 19FT24, 19FT35, 19GT05, 19GT45, 19HT05, 19HT61, 19JT42, 19KT17, 19KT21, 19KT65, 19LT04, 19LT23, 20AT09, 20AT37, 20AT53, 20BT51, 20CT05, 20CT12, 20CT36, 20DT37, 20ET04, 20FT10, 20FT35, 20GT06, KME20H0129, KME20H0906, KME20H1791, KME20H1847, KME20J0848, KME20K1550, KME20L0336, KME20M0492, KME21E1412, KME21G2207, KME21H0839, KME21H0972, KME21H1124, KME21K0660, KME21K1165, KME21L0442, KME21L0443, KME21L1900, KME21L2539, KME21M1391, KME22A1432, KME22B1411, KME22C1068, KME22C2858, KME22D0479, KME22D0482, KME22D0592, KME22D2846 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1847-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.