FDA Device recall Z-1847-2014
Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding
- FDA Device
- Class II
- terminated
- Published 2014-07-02
On 2014-07-02 the FDA classified a Class II recall of Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding by Baxter Englewood, citing baxter Corporation is initiating a field correction following an upgrade to ABACUS Software v3.1, when Clinimix 4.25110 is used in an order, the Nutritional Summary and any labels….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1847-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-05-28 |
| Published | 2014-07-02 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Baxter Englewood |
| Distribution | US |
Product
Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding.
Reason
Baxter Corporation is initiating a field correction following an upgrade to ABACUS Software v3.1, when Clinimix 4.25110 is used in an order, the Nutritional Summary and any labels containing energy information may display 1,000 times less protein than is actually present in the solution.
Identifiers
| code_info | Product Code: 8300-0167, 8300-0168, 8300-0169 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1847-2014%22