FDA Device recall Z-1844-2015
Volcano PV .035 Catheter: Part number: 88901
- FDA Device
- Class II
- terminated
- Published 2015-07-01
On 2015-07-01 the FDA classified a Class II recall of Volcano PV .035 Catheter: Part number: 88901 by Volcano, citing during a relabeling procedure at the manufacturing facility, the products were labeled with a shelf life in accordance with the date of the relabeling and not in accordance with t….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1844-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-06-04 |
| Published | 2015-07-01 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Volcano |
| Distribution | US |
Product
Volcano PV .035 Catheter: Part number: 88901; Product Usage: The Volcano PV .035 is indicated for use in the evaluation of vascular morphology in blood vessels of the peripheral vasculature by providing a cross-sectional image of such vessels. The device is an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures and dimensional measurements from the image.
Reason
During a relabeling procedure at the manufacturing facility, the products were labeled with a shelf life in accordance with the date of the relabeling and not in accordance with the original shelf life date at the time of manufacture
Identifiers
| code_info | Product number 88901; lot number 50002518 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.