RecallRadar

FDA Device recall Z-1840-2022

Roadrunner PC Wire Guide (The Firm LT), Guidewire Reference Part Number/GPN RLPC-35-145 G06866

On 2022-10-05 the FDA classified a Class II recall of Roadrunner PC Wire Guide (The Firm LT), Guidewire Reference Part Number/GPN RLPC-35-145 G06866 by Cook, citing cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices ma….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1840-2022
ClassificationClass II
Statusongoing
Initiated2022-08-16
Published2022-10-05
Last updated2026-09-13 11:56 UTC
CompanyCook
DistributionUS

Product

Roadrunner PC Wire Guide (The Firm LT), Guidewire Reference Part Number/GPN RLPC-35-145 G06866

Reason

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Identifiers

code_infoRLPC-35-145 G06866 UDI-DI: 00827002068667 Lots/Expiration Date: 14828631 06-07-2025 14828643 06-07-2025 14828644 06-07-2025

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1840-2022%22